Damage to the lift may occur.
– The legs must be in the fully closed position before
adjusting the linkage rods.
1.
Close the legs all the way. Refer to 4.2.1 Closing/Opening
Manual Legs, page 17.
Check the squareness of the legs B when in the closed position.
2.
Place a square C on the inside of the legs and base to determine
3.
the 90° alignment D.
Adjust the linkage rods E until 90° alignment is achieved.
4.
Damage to the lift may occur.
– DO NOT adjust the linkage rods so the legs are at
less than 90° alignment.
1078985-G

8.7 Replacing the Electric Actuator

A Bottom nut
B Washer
C Shoulder bolt
D Electric actuator
Maintenance
G Top nut
H Bolt
I Bracket
J Bushing