Accuracy Test; Disposal; Guarantee; Technical Specifications - Microlife BP B2 Basic Manual De Uso

Ocultar thumbs Ver también para BP B2 Basic:
Tabla de contenido
1. Remove the cuff connector 9 from the cuff tube AT and care-
fully pull the bladder through the opening at the edge of the cuff
cover.
2. Hand wash the cuff cover in soapsuds: not hotter than 30 °C.
3. Completely dry the cuff cover by linen drying.
4. Loop the cuff tube back through its opening and carefully place
the bladder flat in the cuff cover.
5. Reattach the cuff connector on the cuff tube.
The bladder must lay straight in the cuff cover, not folded.
Do not use fabric softener.
WARNING: Do not wash the cuff in a washing machine or
dishwasher!
WARNING: Do not dry the cuff cover in a tumble dryer!
WARNING: Under no circumstances may you wash the
inner bladder!

Accuracy test

We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).

Disposal

Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.

9. Guarantee

This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
 Batteries and parts that become worn with use are not included.
 Opening or altering the device invalidates the guarantee.
 The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
 The cuff has a functional guarantee (bladder tightness) for 2
years.
Please contact your local Microlife-Service (see foreword).
6

10. Technical Specifications

Operating condi-
10 - 40 °C / 50 - 104 °F
tions:
15 - 95 % relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
277 g (including batteries)
Dimensions:
131 x 90 x 60.5 mm
Measuring proce-
oscillometric, corresponding to Korotkoff
dure:
method: Phase I systolic, Phase V
diastolic
Measurement range: 20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
4 x 1.5 V alkaline batteries; size AA
Mains adapter DC 6V, 600 mA (optional)
Battery lifetime:
approx. 920 measurements
(using new batteries)
IP Class:
IP20
Reference to
EN 1060-1 /-3 /-4; IEC 60601-1;
standards:
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.