Acoustical Signals - Charger
Ongoing beeps are emitted when the handpiece is in
the charger and the charge contacts are wet.
Troubleshooting
Error
Cause
The 5 sec LED blinks
The residual battery charge
and a short signal is
has dropped below 10%.
produced 5x every
time the button is
pressed or the light
is turned off.
The ongoing exposure The battery lacks sufficient
is interrupted (light
charge.
off signal is emitted)
followed by a 2 sec-
error signal; the device
switches to power-
down mode and resists
further activation.
A 2 sec-error signal
The handpiece has become
is emitted upon
overheated in the course
pressing of the
of use.
Start button.
The handpiece can be used
again once it has cooled down.
Ongoing beeps are
The charge contacts are wet.
emitted and the yellow
LED is flashing fastly
when the handpiece
is put into the charger.
The yellow LED of the
The battery is defective.
charger flashes.
The green LED of
Power outlet carries
the charger fails to
no voltage.
light up, even though
the power cable is
connected to a power
outlet.
Charger is defective.
Solution
Place the handpiece in
the charger and re-charge
the battery.
Place the handpiece in
the charger and re-charge
the battery.
Allow the handpiece to cool
down.
- The handpiece can be
used again, once
pressing the Start button
successfully activates
the light.
Dry the charge contacts.
Do not bend the pins during
drying.
Replace the battery.
Use a different power
outlet.
Have the charger repaired.
Error
Cause
The light intensity is
too low
Maintenance and Care
Replacement of the Battery
Never place the handpiece in the charger without the
battery inserted in the handpiece!
Use 3M ESPE batteries only! The use of other manu-
facturer batteries or non-rechargeable batteries/primary
batteries is a potential hazard and may damage the
device.
Rotate the lid at the lower end of the handpiece
counterclockwise until it hits the stop, and then
remove the lid.
Push a small spatula or similar aid between the
cylinders of the battery until the snap-lock releases
so that the battery protrudes slightly from the hand-
piece body.
Remove the battery from the handpiece.
Push the (replacement) battery into the handpiece
proceeding along the direction shown by the arrow
until the battery locks into place.
Replace the lid and close by rotating clockwise.
Place the handpiece in the charger for 12 hours in
order to fully charge the new battery for the first time.
- The yellow LED of the charger illuminates after
approx. 2 sec; please see also "LED Display of the
Charger".
Solution
Clean the light guide and
the protecting glass in the
light guide mounting hole
(please refer to "Cleaning
the Light Guide").
7