Calculation of term DD
Specified range
Resolution
Accuracy
10% reading + 1 pt
7.3
Power Supply
The equipment power supply is obtained from:
- Rechargeable NiMh batteries - 8 x 1.2 V / 3.5 Ah
- External recharge: 85 to 256 V (50 -60 Hz)
Minimum battery charge life (per NF EN 61557-2)
Test voltage
Nominal charge
Number of measurements
5 s on nominal charge
(with 25 s pause between
each measurement)
Average battery charge life
Supposing a DAR measurement of 1 minute, 10 times per day, with a PI measurement of 10 minutes,
5 times per day, l'autonomie sera d'environ 15 jours ouvrables ou 3 semaines.
Recharge time (C.A 6545 and C.A 6547)
6 hours to recover 100% capacity (10 hours if the battery is completely run down)
0.5 hours to recover 10% capacity (charge life: 2 days approximately)
Note: It is possible to recharge the batteries while performing insulation measurements provided
that the values measured are higher than 20 MΩ. In this case, the recharging time is higher than 6
hours and depends on the frequency of the measurements.
7.4
Environmental parameters
Range of use
-10°C to 40°C, during battery recharging
-10°C to 55°C, during measurement
10% to 80 % RH
Storage:
-40°C to 70°C
10% to 90 % RH
Altitude: < 2000 m
7.5
Construction specifications
Overall dimensions of the unit (L x w x h): 270 x 250 x 180 mm
Weight: approx. 4.3 kg
7.6
Compliance with international standards
Electrical safety as per:
Dual insulation:
Pollution level: 2
Measurement category: III
Max. voltage relative to earth: 1000 V (2500 V in measurement category I)
0,02...50,00
0,01
500 V
1000 V
500 kΩ
1 MΩ
6500
5500
EN 61010-1 (Ed. 2 for 2001), EN 61557 (Ed. 97)
61
2500 V
5000 V
2.5 MΩ
5 MΩ
4000
1500