External Calibration
Set the following parameters:
SETUP - BAL.SCAL.- CAL.Just. - Cal.Ext. (Code 1.1. 9. 1)
The required calibration weight is configured at the factory (see CD-ROM "Specifications"
chapter).
Step
1. Tare balance
2. Start adjustment routine
Once you store the zero point
the display prompts for the required
calibration weight (flashing display)
3. Apply the prompted calibration
weight (in this example 5000 g).
Weight too low:
a minus sign "-" is shown
Weight too high:
a plus sign "+" is shown
The display stops flashing as soon as
the weight value is within the defined
limit.
4. Calibration/adjustment executed;
then the calibration weight
is displayed
5. Remove the calibration weight
16
Key (or instruction)
w
Q
Display
g
0.0
Cal.Ext.
- 5000.0
g
g
5000.0
Cal.end
+ 5000.0
g
g
0.0