8) During nebulizing, the maximum permissible tilting angle of the device
is 45° (from the vertical line). If the angle is exceeded, the liquid may spill
from the metal mesh in nebulizing head.
9) If an operating device is tilted to the extent that the liquid is not in
contact with the nebulizing head, the device will shutdown after a delay
time of 10 seconds (depending on the liquid type).
10) When there is not much liquid left, you may tilt the device towards
yourself to allow liquid gathers at the nebulizing head side in order to
make full use of the residual liquid. During operation be cautious not to
shake the device because it may cause device malfunction, such as
automatic shutdown
7. After completion of treatment, shut it down by pressing the ON/OFF
switch; if power adapter is connected, you are also required to unplug it.
8. Remove the mouthpiece/mask and the connector in an order reverse to
the corresponding installation order.
9. Remove the outer cover.
10. Remove the nebulizer: hold the main
unit with left hand, press the PRESS-key
(over the jack for adapter) with your
left thumb along the direction of arrow
(see Fig. 13); hold both sides of the lower
end of nebulizer the right hand to gently
pull it out.
11. Remove the residual liquid to empty the nebulizer
and clean the device. For details please refer
to the section – Cleaning and Disinfecting.
Note: You are advised to, after each use, allow the device
running with clean water instead of liquid for 1-2
minutes in order to clean the nebulizing head.