Formula Editor - Xylem SI Analytics TitroLine 7500KF Manual De Instrucciones

Ocultar thumbs Ver también para SI Analytics TitroLine 7500KF:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

150
Example: The blank value of a chloride titration is defined with the support of an extra method. The result in ml is
thereby automatically written into global memory M01 by using the name "Blanc value" (Fig. 96). The blank value
is then automatically deducted from the titrant consumption within the chloride method.
Fig. 96
The menu for the global memory can always be accessed by pressing <Shift> or via system settings. The name
or values can be changed by using <EDIT> and have the methods shown that are used in the global memories
(Fig. 97).
Fig. 97

Formula Editor

4.6.4
The formula editor is intended to complement the existing standard formulas. The standard formulas are loaded
and can then be modified. The original standard formula itself is never changed.
4.6.4.1 Start and Work with the Formula Editor
With <EDIT> you are going to «Edit method», «New method» or «Default methods» and select then «Result»
(Fig. 98).
Fig. 98
Tabla de contenido
loading

Tabla de contenido