Xylem Si Analytics ViscoClock Manual De Instrucciones página 28

Aparato de medicón de la viscosidad
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

26
3.4 Insertion of the Viscosity Measuring Unit ViscoClock into the thermostat bath
The transparent thermostat has to be equipped with a VZ 5402 manual measuring insert in order to ac-
commodate the Viscosity Measuring Unit ViscoClock.
• The Viscosity Measuring Unit ViscoClock, with filled viscometer, is inserted into the manual measuring
insert {22) of the thermostat bath.
Fig. 5
3.5 Connecting the Viscosity Measuring Unit ViscoClock
CAUTION
Prior to plugging the power supply (29) into the socket, make sure that the mains-unit voltage matches the
mains voltage.
• Plug the 9 V/ DC jack plug (19) into the back panel of the electronic measuring unit
• Plug the TZ 1858 universal voltage power supply for 100 - 240 V~, 50/60Hz, into the mains socket.
3.6 Connecting the Viscosity Measuring Unit ViscoClock to a computer (PC) or
printer
There is an RS-232-C Interface (18) on the rear side of the electronic measuring unit (13).
• lf desired, connect a computer (PC) or printer for documentation purposes.
Plug connection: 4-channel mini DIN round plug.
Required data cable, which is not included in the scope of delivery, → accessories, chapter 5.
Fig. 6
The transfer parameters of the RS-232-C Interface
are firmly set:
• - 4800 baud
• - 7 bit word length
• - 2 stop bits
• - no parity
Initial operation
Thermostat bath
surface
Tabla de contenido
loading

Tabla de contenido