9.2 Electromagnetic immunity

Guidance and Manufacturer's Declaration – Electromagnetic Immunity
ARTROMOT®-S3 is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that the ARTROMOT®-S3 unit is
used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic
± 6 kV
discharge (ESD) to
contact
IEC 61 000-4-2
± 8 kV air
Electrical fast
± 2 kV for
transient/bursts
power supply lines
to
± 1 kV for input/
IEC 61 000-4-4
output lines
Surges to
± 1 kV
IEC 61 000-4-5
line-to-line voltage
± 2 kV
line-to-earth
voltage
Voltage dips, short
< 5 % U
interruptions and
(> 95 % dip in U
voltage variations
for 1/2 cycle
on power supply
40 % U
input lines to
(60 % dip in U
IEC 61 000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
< 5 % U
(> 95 % dip in U
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field to
IEC 61 000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
t
Compliance level
± 6 kV
contact
± 8 kV air
± 2 kV for
power supply lines
± 1 kV for input/
output lines
± 1 kV
line-to-line voltage
± 2 kV
line-to-earth
voltage
< 5 % U
t
t
)
(> 95 % dip in U
t
for 1/2 cycle
40 % U
t
t
)
(60 % dip in U
t
for 5 cycles
70 % U
t
t
)
(30 % dip in U
t
for 25 cycles
< 5 % U
t
t
)
(> 95 % dip in U
t
for 5 s
3 A/m
80
Electromagnetic
environment - guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power should be that of
a typical commercial or hospi-
tal environment.
Mains power should be that of
a typical commercial or hospi-
tal environment.
Mains power should be that of
)
a typical commercial or hospital
t
environment.
If the user of the
)
ARTROMOT®-S3 unit requires
t
continued operation during
power mains interruptions,
)
it is recommended that the
t
ARTROMOT®-S3 unit be
powered from an uninter-
)
ruptible power supply or a
t
battery.
Power frequency magnetic
fields should be at levels
characteristics of a typical
location in a typical commer-
cial or hospital environment.