5.6 Examples of use/
sample protocols
5.6.1 Isolated
adduction/abduction
1. Complete the mechanical settings and
adjust the carriage to the patient as
described in section 4.2.
2. Press the MENU key on the program-
ming unit to access level 1 (M1).
3. Press the parameter key for internal
rotation
or external rotation
and press the +/- keys to set the
carriage to the desired rotation position
to be maintained during the isolated
abduction/adduction movements.
4. Next set the desired range of motion
for adduction
and abduction by
first pressing the parameter key for
adduction and set the values with the
+/- keys.
5. Set the abduction value in the same
way
.
Note!
– For an exclusive adduction/abduction
protocol, motor B for the rotation
movement must be turned off as
described in steps 6 to 8 below.
– Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the MENU key on the program-
ming unit to access level 3 (M3).
7. Press the Motor B ON/OFF parameter
key
, to activate the parameter.
8. Deactivate motor B by pressing the
parameter key Motor B ON/OFF again
or the „–" (minus) key. The check mark in
the circle next to the symbol must have
disappeared.
9. Save the settings with the STOP key, set
the carriage to the starting.
5.6.2 Isolated
internal/external rotation
1. Complete the mechanical settings and
adjust the carriage to the patient as
described in section 4.2.
2. Press the MENU key on the program-
ming unit to access level 1 (M1).
3. Press the parameter key for adduction
or abduction
the +/- keys to set the carriage to the
desired position to be maintained during
the isolated rotation movements.
4. Next set the desired range of motion
for internal/external rotation by first
pressing the parameter key for internal
rotation
and set the values with
the +/-keys.
5. Set the external rotation value in the
same way
Note!
– For an exclusive rotation protocol,
motor A for adduction/ abduction
must be turned off as described in
steps 6 to 8 below.
– Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the MENU key on the program-
ming unit to access level 3 (M3).
7. Press the Motor A ON/OFF
parameter key to activate the parameter.
8. Deactivate motor A by pressing the
parameter key Motor A ON/OFF again
or the „–" (minus) key. The check mark in
the circle next to the symbol must have
disappeared.
9. Save the settings with the STOP key,
set the carriage to the starting position
with the START key and initiate therapy
by pressing the START key again.
71
and press
.
,