F1.4.1 – Automatic zero setting
OFF
Automatic zero setting disabled (not available in
NTEP mode)
0.5 d Automatic zero setting within +/–0.5 d
1 d
Automatic zero setting within +/–1.0 d
(not available in OIML mode)
3 d
Automatic zero setting within +/–3 d (not available
in OIML mode)
F1.4.2 – Power up zero
OFF
Power up zero disabled
2
Power up zero within +/–2 %
Power up zero within +/–10 %
10
20
Power up zero within +/–20 %
F1.4.3 – Pushbutton zero
0
Pushbutton zero disabled
2
Pushbutton zero with
+/–2 % zero setting range
10
Pushbutton zero with
+/–10 % zero setting range
20
Pushbutton zero with
+/–20 % zero setting range
F1.5.1 – Automatic taring
On
Automatic taring enabled
OFF
Automatic taring disabled
F1.5.2 – Auto clear tare
On
Clearing tare automatically enabled
Clearing tare automatically disabled
OFF
F1.5.3 – Tare Interlock
On
The forks must be unloaded to zero before the tare
weight can be cleared.
OFF
Function disabled
F1.5.4 – Auto tare threshold
This menu item is only available if F1.5.1=On is set.