Info:
If the application period is exceeded significantly, the plastic parts may
take on the smell of the disinfectant.
Rinse all parts thoroughly in running water.
CAUTION: Risk of an allergic reaction
Residues of the disinfectant solution can cause aller-
gic reactions or inflammation of the mucous mem-
brane. Therefore, all chemically disinfected parts must
be rinsed off thoroughly.
Dispose of the used solution (the diluted solution can be dis-
posed of down the drain).
5.5 Drying
After disinfection, place the individual components of the
®
INQUA
nebuliser on a dry, clean and absorbent surface.
You can remove excess water more quickly by shaking all
parts.
Let them dry completely.
The components should not be dried in damp rooms (e.g., the
bathroom).
64
Prepare a 2% solution of
®
Bomix
plus by mixing 10 ml of
the concentrate with 500 ml tap
water.
Place all the individual parts in
the prepared solution and leave
them to soak for 5 minutes.