6 C L E A N I N G / H Y G I E N E
X When cleaning the Cooling Unit Cap
in isopropyl alcohol, remove the cap
lock by pulling it up on one side.
X First press the cap lock on one side
and pull it up on the opposite side.
X Then slide the cap lock on the raised
side towards the opposite side.
X Unscrew the Cooling Unit base
from the Vaporizer by turning it 90°
counterclockwise.
X Lever out the large Base Seal Ring
with Filling Chamber Tool at the
recess in the groove.
X Remove the small Base Seal Ring.
X Press the Screen with the Filling
Chamber Tool from the top to the
bottom out of the Cooling Unit base.
73