Vapormed Mighty+ Medic Modo De Empleo página 251

Ocultar thumbs Ver también para Mighty+ Medic:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 125
INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MIGHTY + MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the
MIGHTY + MEDIC Vaporizer should assure that it is used in such an environ-
ment.
Immunity
test
Conducted RF
IEC 61000-4-6
IEC 60601
test level
3 Vrms
150 kHz to 80
MHz
6 Vrms
ISM/Amateur
inside 150 kHz
Radio bands
inside 150 kHz
to 80 MHz
3 V/m
80 kHz to 2.7
GHz
Compliance
level
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the MIGHTY +
MEDIC Vaporizer,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation
distance:
3 Vrms
150 kHz to 80
MHz
6 Vrms ISM
Radio bands
to 80 MHz
3 V/m
80 kHz to 2.7
GHz
A-5
Electromagnetic
environment –
guidance
d = 1.2 √P
d = 1.2 √P
80 MHz to
800 MHz
loading