INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
Immunity
test
Radiated RF
IEC 61000-4-3
Note 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagne-
tic propagation is affected by absorption and reflection from
structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio
a)
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the
MIGHTY + MEDIC Vaporizer is used exceeds the applicable RF compliance
level above, the MIGHTY + MEDIC Vaporizer should be observed to verify
normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the
MIGHTY + MEDIC Vaporizer.
Over the frequency range 150 kHz to 80 MHz, field strengths should be
b)
less than 3 V/m.
IEC 60601
test level
RF
communica-
tion equipment
inside 80 MHz
to 6 GHz
Compliance
level
RF
communica-
tion
equipment
inside 80
MHz to 6 GHz
Where P is the maximum
output power rating of the
transmitter in watts [W]
according to the
transmitter manufacturer
and d is the recommended
separation distance in
meters [m].
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
survey
than the compliance level
in each frequency range
A-6
Electromagnetic
environment –
guidance
d = 2.3 √P
800 MHz to
2.7 GHz
, should be less
a)
.
b)
loading