INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
2 Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The MIGHTY + MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the
MIGHTY + MEDIC Vaporizer should assure that it is used in such an envi-
ronment.
Immunity
test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical
fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601
test level
± 8 kV contact
± 8 kV contact
discharge
discharge
± 15 kV air
± 15 kV air
discharge
discharge
± 2 kV for
± 2 kV for
power supply
power supply
lines
lines
± 1 kV for
± 1 kV for
input/output
input/output
lines
lines is not
applicable
± 1 kV line(s)
± 1 kV line(s) to
to line(s)
line(s)
± 2 kV line(s)
± 2 kV line(s) to
to earth
earth
Compliance
level
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Mains power quality
should be that of a
typical residential,
commercial or
hospital environment.
Mains power quality
should be that of a
typical residential,
commercial or
hospital environment.
A-3
Electromagnetic
environment –
guidance