5.9
Functions of the Remote Interface
The remote interface provides two inputs (START and RESET) and three outputs
(TESTING, PASS and FAIL).
RESET:
START:
TESTING 1,
TESTING 2:
PASS 1,
PASS 2:
FAIL 1,
FAIL 2:
��������
�����������
The function buttons on the front panel are disabled when any two pins on the rear
panel for Reset, Start or Common are shorted.
Please check the remote control device.
Hipot Electrical Safety Testers Models H111 and H216
���������
���������
������
������
Connecting "RESET" (pin 1) to "COMMON" (pin 3) will reset the
tester (provides the same functionality as the RESET button on
the front panel).
In status READY mode, connecting "START" (pin 2) to
"COMMON" (pin 3) will start the test (provides the same func-
tionality as the TEST button on the front panel).
During test, the "TESTING 1" (pin 4) and TESTING 2"
(pin 5) output pins will be shorted.
If the test passed, the "PASS 1" (pin 6) and "PASS 2"
(pin 7) output pins will be shorted.
If the test failed, the "FAIL 1" (pin 8) and "FAIL 2" (pin 9)
output pins will be shorted.
������ �
������� �
�������
����������
���������
������ �
������� �
�������
�������
������
�����
�����
������
������
�������
�������
����������
��������
��������
��������
��������
��������
��������
��������
��������
��������