Examples:
*IDN?
Returns AEMC, HT216, 0, FW:1.00
*OPC
Function:
The command form (*OPC) sets the operation complete bit (bit 0) in the Standard
Event Status Register (SESR) when all pending operations are finished.
The query form (*OPC?) tells the programmable EST to place an ASCII 1 in the Output
Queue when the power supply completes all pending operations.
Syntax:
*OPC
*OPC?
Returns:
1
*RST (no query form)
Function:
Set all control settings of EST to their default values but does not purge stored set-
ting. The equivalent panel control will be set as below:
TEST
Group and Step
Syntax:
*RST
*SRE
Function:
Set the contents of the Service Request Enable Register (SRER). The query form
returns the contents of the SRER. Bit 6 of the SRER is always zero. The bits on
the SRER correspond to the bits on the SBR.
Syntax:
*SRE <NR1>
*SRE?
Returns:
<NR1>
is in the range from 0 through 255.
22
Front Panel Control
Hipot Electrical Safety Testers Models H111 and H216
Default Setting
OFF
0:0