5.6.2 Table of Utility Functions
To select a function to program while in the Utility mode, rotate the knob until the
desired function appears in the display.
Function
TEST MODE
ARC MODE
ARC CURRENT
AC (TEST)
FREQUENCY
TEST
CONTROL
MODE
DATA LOCK
CONT.
CALIBRATION
IR TEST MODE
INTERFACE
FAIL SETTING
Hipot Electrical Safety Testers Models H111 and H216
Option
From
STEP 1
From the
present step
DISABLE
ENABLE &
STOP
ENABLE &
CONTINUE
50Hz
60Hz
MODE 1
MODE 2
MODE 3
MODE 4
UNLOCKED
LOCKED
Stop on Fail
Stop on Pass
Timer
RS-232 Baud
Rate
GPIB Address 00 to 31
Stop
Continue
Description
The test mode procedure always begins
from step 1 to the end of the group.
e.g. 3:1 - 3:6, 4:1 - 4:6
The test mode procedure always begins with
the selected step to the end of the group.
e.g. 3:3 - 3:6, 4:3 - 4:6
Disables the ARC detection function
Enables ARC detection and stops the test
when ARC is detected
Enables ARC detection and continues the
test when the arc is detected
Sets the current level of ARC detection
(1 to 40mA)
Sets AC Hipot output frequency to 50Hz
Sets AC Hipot output frequency to 60Hz
Control mode from the front panel.
Mode 1: Reset first (recommended)
(press RESET button before test)
Mode 2: Press START button directly
Mode 3: REMOTE I/O is enabled
(the start button is disabled)
Mode 4: Reserved.
Accepts and saves all parameters from the
test/step and utility
Locks out any changes to parameters from
the test/step and utility until unlocked
Calibrates the short circuit resistance of test
leads for test of continuity check.
Stop on fail at the DUT detection
Stop on pass at the DUT detection
Reach the time of determination for pass
or fail
1200, 2400, 4800, 9600
The operation will stop upon the failure
occurred at any step of group test
The operation will not stop until all 16 group
steps have been tested
49