3.1
Electrical Specifications
3.1.1 AC Specifications
AC Voltage Range:
Voltage Step:
Voltage Regulation:
Voltage Accuracy:
Max Current*:
Current Limit:
Current Accuracy:
3.1.2 DC Specifications (H216 Only)
DC Voltage Range:
Voltage Step:
Voltage Regulation:
Voltage Accuracy:
Max Current*:
Current Limit:
Current Accuracy:
*Note: The Hipot testers are not designed for continuous use at the maxi-
mum rating. The test procedure will automatically stop when temperature
overload is detected.
Stop the instrument for 10 minutes when continuously proceeding 30 to
40mA withstanding test for one hour.
10
C
HAPTER
S
PECIFICATIONS
100V to 5000V (5kV)
5V/step
1% of Reading ± 5V
2% of Reading ± 10V
30 to 40mA (above 500V; 180s max test time)
0.10 to 29.99mA (above 500V; continuous test)
0.10 to 10mA (below 500V; continuous test)
0.10 to 40mA, 0.02mA/step
2% of Reading ± 2cts
100V to 6000V (6kV)
5V/step
1% of Reading ± 5V
2% of Reading ± 10V
0.10 to 10mA (above 500V; continuous test)
0.10 to 2mA (below 500V; continuous test)
0.10 to 10mA, 0.01mA/step
2% of Reading ± 2cts