Warning: Don't near active HF surgical equipment and the RF shielded
room of an ME system for magnetic resonance imaging, where the inten-
sity of EM disturbances is high.
Warning: Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer
than 30 cm to any part of this appliance (HNS018839), including power
cable. Otherwise, degradation of the performance of this equipment could
result.
If the appliance does not pump the air in cuff normally or it is not working in
alternating inflating and deflating due to the electromagnetic disturbances,
do not use it and try to place the appliance in other location.
For maintaining the basic safety and essential performance with regard to
electromagnetic disturbances for the expected service life, please precau-
tion to take if the use location is near (e.g. less than 1,5 km from) from AM,
FM or TV broadcast antennas.
EMC Information
The appliance is complied with the following emission and immunity standard,
and test level:
Emissions Test Standard
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emis-
sions
IEC 61000-3-3
Immunity Test standard
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
40
Compliance
Group 1
Class B
Class A
Comply
IEC 60601-1-2 Test Level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 2 kV, for power supply
lines
± 1 kV signal input / output
100 kHz repetition
frequency
Compliance Level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 2 kV, for power supply
lines
± 1 kV signal input / output
100 kHz repetition
frequency
loading