Fault/error code
Workflow LED does not light
up
Workflow LED illuminated red
001-288XX-XXX
002-288XX-XXX
004-288XX-XXX
005-288XX-XXX
006-288XX-XXX
010-288XX-XXX
019-288XX-XXX
136 • seca 651/650
Cause
Measurement not yet started
No network connection
WiFi function deactivated
Workflow LED defective
ID not found in EMR System or in the
seca software
Device has not saved measured results
to the clipboard
Measured results have not been sent to
an EMR System or to seca software
No connection to the server
No data transmission possible
Barcode scanned at the wrong time
Barcode invalid or damaged
Patient ID/user ID not found
Unable to transmit measurement to
EMR System
Update failed
Overcurrent at USB interface of weigh-
ing platform
Remedy
Start measurement
(è Starting the mea-
surement
procedure, page 102)
Set up network connection
network functions, page 123
Activate the WiFi function
activating the WiFi function, page 125
Inform seca Service
Create the ID in the EMR System or in the
seca software
Repeat measurement
• Repeat measurement
• Check WiFi connection
• Check LAN cable
• Check network settings
network functions, page 123
• Check network settings
network functions, page 123
• Check Workflow settings for the
seca connect 103
or
seca analytics 125
software
Repeat measurement and maintain correct
operating sequence
è Restarting the device, page 137
• If the error recurs, inform seca Service
Use a valid, undamaged barcode
In the database (
seca analytics 125
EMR System):
• Create patient file
• Create user account
Check Workflow settings for the
or
seca connect 103
seca analytics 125
software
• Restart update
• If the error recurs, inform seca Service
• Check connected USB device
• If the error recurs, do not use the USB
device
è Setting up
è Activating/de-
è Setting up
è Setting up
or
loading