Accuracy for Home Use by Lay-Users
®
The Embrace
WAVE+ Blood Glucose Meter result may vary slightly from your actual blood
glucose value. This may be due to slight differences in technique and the natural variation
in the test technology.
The chart below shows the results of a study where total of 143 fresh blood samples from
fingertip, palm and forearm were collected by lay users. 16 samples were below 75mg/dL.
127 samples were above 75mg/dL. The following accuracy results were obtained.
Sample Source
Lay users Fingertip vs. YSI
Lay users Palm vs. YSI
Lay users Forearm vs. YSI
76
Table 1: Accuracy for blood glucose level <75mg/dL
Within ± 5mg/dL
12/16 (75%)
15/16 (94%)
13/16 (81%)
Within ± 10mg/dL Within ± 15mg/dL
16/16 (100%)
16/16 (100%)
16/16 (100%)
16/16 (100%)
16/16 (100%)
16/16 (100%)