CAUTION:
• Always check the expiration date. DO NOT use control solutions if they are expired.
• Mark the newly opened bottle of control solution with the date opened. Discard any
unused control solution three months after opening.
• For accurate test results, allow the control solution to adjust to its surroundings for at
least 30 minutes before running the control test:
▪ Temperature: 41 to 113 °F;
▪ Relative Humidity: 20 to 90% RH.
• DO NOT FREEZE. Store the control solutions at 39˚F - 86˚F.
• Do not drink the control solution. It is not intended for human consumption.
• Avoid contact of the solution with the skin and the eyes as this could cause
inflammation.
• Discard used control solution bottles according to your local regulations.
Performing a Control Solution Test
Materials you will need:
• Embrace
®
WAVE Glucose Control Solution Low (L1) and High (L2)
• Your Embrace
• A new Embrace
32
®
WAVE+ Blood Glucose Meter
®
WAVE Blood Glucose Test Strip