2. Discard the first drop of control solution onto a clean, unused tissue or
paper towel, then wipe the bottle tip clean to ensure that a proper sample
is applied to the test strip.
3. Dispense a second drop of control solution Low (L1) or High (L2) onto a
clean, hard, dry nonabsorbent surface to avoid cross contamination. Do
not apply control solution to the test strip directly from the bottle.
4. Replace the bottle cap on the control solution bottle immediately after use.
NOTE:
• The test will not start if you apply your control solution not directly to the edge of the
sampling end of the test strip. The test starts when the meter detects the control
solution. During the test the meter counts down from 5 to 1.
• Tightly close the control solution vial.
• Do not touch the test strip once the meter has started the countdown.