7.3 Instruction Set for RS-Communication

The commands consist of three parts: Address
Every command must be completed with the ASCII - sign <CR> and <LF> (Carriage Return and Line Feed).
Only if the respective action has ended the answers will be returned to the computer.
Example:
The command to dose 12.5 ml shall be sent to the burette TITRONIC
The command consists of the characters: 02DA12.5<CR LF>
In detail:
02
DA
12.5
<CR LF> = Control character as command end
Command
Description
aaAA
automatic allocation of device address
aaMC1...XX
choosing a method
aaBF
„filling burette". Aufsatz wird gefüllt.
aaBV
output of dosed volume in ml
aaDA
dose volume without filling, with adding the volume
aaDB
dose volume without filling, reset of the volume
aaDO
dose volume with filling, without adding the volume
aaGDM
dosing speed in ml/min
aaGF
filling time in seconds (min is 20, default 30)
aaES
„ESC" function one step backwards
aaEX
„EXIT" function.back to main menu
aaGDM
dosing speed in ml/min (0.01 – 100 ml/min)
aaGF
filling time in sec (adjustable 20 – 999 seconds)
aaGS
output serial no. Of device
aaLR
output report (short report)
aaLI
output method content
aaLO
output documentation (as configured)
aaRH
request of identification
aaRC
send last command
aaRS
report status
possible answers are:
„STATUS:READY" for ready
„STATUS:dosing" dosing
„STATUS:filling" filling
„ERROR:busy" if no interchangeable unit has been attached
aaSM
start selected method
aaSEEPROM
EEPROM reset to factory defaults
aaSR
stop the actual function
aaVE
Version number of the software
Command
Variable, if necessary
and end of command
= Device address
= Dosage command with filling and zero points of the display
= Volume in ml to be dosed
two-digit aa, e.g.: 01
e.g.: DA
e.g.: 14
<CR> <LF>
®
300 with the address 2.
Reply
aaY
aaY
aaY
aa0.200
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaGS08154711
aaY
aaIdent: TITRONIC 300
aa"last command"
aaStatus:"text
aaY
aaY
aaY
aaVersion:
Tabla de contenido