Frequency measurement (electronic)
Range
40,000Hz
400,00Hz
4,0000kHz
40,000kHz
400,00kHz
4,0000MHz
40,000MHz
100,00MHz
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and
<100kHz; 5Vrms min @ 20% to 80% duty cycle and > 100kHz
Frequency measurement (electrical)
Range
40,00Hz ...
10,000kHz
Sensitivity: 1Vrms
Duty cycle measurement
Range
0,10 ... 99,0%
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz
Temperature measurement
Range
-50.0...1200,0C
-58.0...2192,0F
* probe accuracy not included
70
Resolution
0,001Hz
0,01Hz
0,0001kHz
0,001kHz
0,01kHz
0,0001MHz
0,001MHz
0,01MHz
Resolution
0,01Hz ... 0,001kHz
Resolution
0,01%
Resolution
0,1C
0,1F
CMM-40 – USER MANUAL
Basic uncertainty
(0,1 % m.v. + 1 digit)
Not specified
Basic uncertainty
0,5 % m.v.
Basic uncertainty
(1,2 % m.v. + 2 digits)
Basic uncertainty
± (1% m.v. + 2,5C)