6. Commissioning, operation
6. Commissioning, operation
Commissioning may only be carried out by trained and qualified personnel. The calibra-
tion system was comprehensively checked ex works for tightness.
GB
6.1 Switching on and off
To switch on the calibration system,
press the Power key (1). Press the
Power key again to switch off the
calibration system.
After switching on the version of the
firmware is indicated on the display for
approx. 1 second.
6.2 Zero point adjustment
To set the displayed value as zero
point, simply press the ZERO key (3).
The display changes immediately to the
0 value.
6.3 MIN/MAX value
The digital pressure gauge saves the
minimum and maximum pressure.
If the MAX/MIN key (4) is pressed once,
the maximum pressure from the memory is displayed. If the MAX/MIN key is pressed
again, the minimum pressure is displayed from the memory.
After 2 seconds, the display switches back to the operating mode. To clear the MAX/MIN
memory, press the MAX/MIN key (4) for at least 2 seconds until the message "CLr" is
displayed.
The bargraph display at the bottom of the display indicates the applied pressure relative
to the total measuring range.
14
WARNING!
Only connect or disconnect test and calibration installations once the
system has been depressurised!
Due to a possible Tare setting, the displayed pressure value can differ
from the applied pressure.