WARNING
Risk of patient injury due to inadvertent activation of the
ligation instrument!
Never use the AUTOSTART function.
Do not switch on the HF current before the active
electrodes are in contact with the tissue to be
coagulated.
1.
Place the jaws A on the operating site.
2.
Place the tissue to be sealed between the electrodes of
the jaws A.
3.
Close the jaws A to grasp the tissue.
The tissue is grasped.
4.
Use the two-position ratchet of the handle B to adjust
the pressure on the tissue to best match the amount of
grasped tissue.
The tissue is clamped.
5.
Using the foot switch of the HF device, activate the HF
current for coagulation:
A continuous acoustic signal sounds during
the entire sealing process to indicate that
power is being supplied.
An alternating acoustic signal indicates the
end of the sealing process.
6.
Release the foot switch to stop the supply of power.
7.
Using the handle B, open the jaws A.
The tissue is coagulated.
TissueSeal / TissueSeal PLUS
5 Operation