1.
To Select CPAP or APAP Mode (for XT Auto)
1.
Press "MENU" to select [Mode
2.
Hold the "UP" and "DOWN" button, and then simultaneously press the "MENU"
button for one second. Meanwhile, the LCD screen [Mode
blinking to allow you to switch the mode from CPAP to APAP.
3.
Press "UP" or "DOWN" button to select CPAP or APAP mode.
4.
After selecting the mode, press "MENU" to confirm.
5.
Press "START/STANDBY" button to go back to standby screen or leave the device
to automatically go back to standby screen 20 seconds later.
2.
To Set the Therapy Pressure
1.
Press "MENU" to select [P
XX.XcmH
2.
Hold the "UP" and "DOWN" button, and then simultaneously press the "MENU"
button for one second. Meanwhile, the LCD screen [P
blinking to allow you to adjust the therapy pressure from 4 to 20 cm H
3.
Press "UP" or "DOWN" button to increase or decrease the pressure setting. The
increment or decrement of pressure setting is 0.5 cm H
4.
After selecting the prescribed pressure, press "MENU" to confirm.
5.
Press "START/STANDBY" button to go back to standby screen or leave the device
to automatically go back to standby screen 20 seconds later.
6.
In APAP mode, follow same instructions to set the initial pressure (P
pressure (P
), and minimum pressure (P
max
NOTE–If you try to set an maximum pressure lower than P
P
will automatically to be set lower than P
min
3. To Enable/Disable PVA (Pressure Variation Algorithm) (for XT Prime / XT Auto)
1.
Press "MENU" to select [c on/off] menu while in the standby screen.
2.
Hold the "UP" and "DOWN" button, and then simultaneously press the "MENU"
button for one second. Meanwhile, the LCD screen [c off] should start blinking to
allow you to disable or enable PVA function.
3.
Press "UP" or "DOWN" button to switch "on" or "off" the setting.
4.
After selecting the setting, press "MENU" to confirm.
5.
Press "START/STANDBY" button to go back to standby screen or leave the device
to automatically go back to standby screen 20 seconds later.
4.
To Enable/Disable Low Pressure Auto Off (for XT Prime / XT Auto)
1.
Press "MENU" to select [Low P on/off] menu while in the standby screen.
2.
Hold the "UP" and "DOWN" button, and then simultaneously press the "MENU"
button for one second. Meanwhile, the LCD screen [Low P on] should start blinking
to allow you to disable or enable low pressure auto off function.
3.
Press "UP" or "DOWN" button to switch "on" or "off" the setting.
4.
After selecting the setting, press "MENU" to confirm.
5.
Press "START/STANDBY" button to go back to standby screen or leave the device
to automatically go back to standby screen 20 seconds later.
CPAP] menu while in the standby screen.
CPAP] should start
O] menu while in the standby screen.
2
XX.XcmH
O] should start
2
O.
2
O.
2
), maximum