TROUBLESHOOTING
NOTE– Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO THE HOME MEDICAL
EQUIPMENT PROVIDER WHERE YOU PURCHASED IT, please take a few moments to check for these possible causes:
PROBLEM
Unit does not turn on, but green external power light is
illuminated.
Pump runs, but there is no vacuum.
Low vacuum.
Battery will not charge but charge light is illuminated.
SPECIFICATIONS/CLASSIFICATIONS
Size
7305P & 7305D Series
Weight
7305D Series
7305P Series
Electrical Requirements
Internal Rechargeable Battery
7305P Series
7305D Series
Vacuum Range
(NOTE– Using Flovac Liners may impair the device's performance.)
7305P & 7305D Series
Air Flow @ pump inlet:
Container Capacity
7305 D & P Series
7305 D & P Series - Optional
Environmental Conditions
Operating Temperature Range
Operating Relative Humidity
Operating Atmospheric Pressure
Storage & Transport Temperature Range
Storage & Transport Relative Humidity
Storage & Transport Atmospheric Pressure
Warranty
7305P Series
7305D Series
Internal Battery
Approvals
7305D & 7305P Series
7305P meets RTCA/D-160D
Equipment Classifications
With respect to protection from electric shock
Degree of protection against electric shock
Degree of protection against ingress of liquids
Mode of Operation
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
ISO Classification
7305P Series only - Electrically powered medical suction equipment for field and transport use according to ISO 10079-1 : 1999
High Flow/High Vacuum
7305D Series - Electrically powered medical suction equipment for non-transport use according to ISO 10079-1 : 1999
EN - 8
ACTION
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
3. If running from an internal battery, ensure that your unit has a battery installed.
4. If battery is installed, check that it is fully charged.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off is not activated.
4. Check for leaks or cracks in container assembly.
1. Use vacuum adjustment knob to increase vacuum level.
2. Check system for leaks.
3. Push vacuum adjustment knob in toward unit and then release.
1. Ensure that unit is equipped with an internal battery by contacting your DeVilbiss equipment provider.
2. Verify that charge light turns on.
3. Check electrical connections during charging.
4. Ensure wall outlet is live by plugging in a lamp.
H x W x D inches (cm)
9.0 x 7.0 x 8.0 (22.9 x 17.8 x 20.3)
lb. (kg)
3.8 (1.7)
6.3 (2.9)
100-240 V AC 50/60 Hz .75 A max; 12 V DC, 33 W max
Factory Equipped
Not Factory Equipped (provider-installed option)
80 to 550 mm Hg
27 LPM (free flow) typical (may be less when running from internal battery)
800 ml (cc) Disposable
1200 ml (cc) Reusable
32°F (0°C) - 104°F (40°C)
0-95%
10.2 psi (70 kPa) - 15.4 psi (106 kPa)
-40°F (-40°C) - 158°F (70°C)
0-95%
7.3 psi (50 kPa) - 15.4 psi (106 kPa)
Two-years limited, excluding internal battery and container
Two-years limited, excluding container
90-day
IEC 601-1; CAN/CSA-C22.2 No. 601.1-M90; UL 60601-1, EN 60601-1-2
DO-160D - section 21 Category M For battery operation only Airline use
Class I and internally powered
Type BF Applied Parts
IPX2 and ordinary power supply
Intermittent Operation: 30 minutes on, 30 minutes off
SE-704