Ottobock C-Leg 3C98-3 Instrucciones De Uso página 82

Ocultar thumbs Ver también para C-Leg 3C98-3:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 177
Beep signal
30 x long
Error while charging the product
LED on
LED on bat­
power
tery charger
supply
82
Vibration signal
1 x long, 1 x short
repeated every
3 seconds
Continuous
Error
Country-specific plug adapter not
fully engaged on power supply
Non-functional socket
Defective power supply
No connection between battery
charger and power supply
Defective battery charger
Event
Severe error/indic­
ation of safety
mode activation
e.g. one or more
sensors are not oper­
ational.
Total failure
Electronic control no
longer possible.
Safety mode active or
undetermined valve
state. Undetermined
product behaviour.
Resolution
Check whether the country-spe­
cific plug adapter is fully
engaged on the power supply.
Check socket with another elec­
tric device.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Check whether the charging
cable plug is fully engaged on the
battery charger.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Required action
Walking possible with
restrictions. Please note
the possible change in
flexion/extension resist­
ance.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger. The battery charger
must be connected for at
least 5 seconds before it
is disconnected.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.
Tabla de contenido
loading

Este manual también es adecuado para:

C-leg 3c88-3

Tabla de contenido