- The screen displays the message LO : The analysed temperature is
below the measuring range of the selected mode, i.e. less than 32°C
(89.6°F) in BODY mode and less than 0°C (32°F) in SURFACE mode.
This message displays in various cases - below are a list of the main
cases:
CAUSES OF LO MESSAGE
DISPLAY
Temperature reading hampered
by hair, sweat,...
Temperature reading hampered
by an air flow...
Temperature reading too close
together, ThermoFlash® has not
had time to reset.
The measuring distance is too
great.
All manuals and user guides at all-guides.com
SOLUTIONS
Make sure there is no
obstruction prior to taking the
temperature.
Be careful not to be in the axis
of an airflow, this could cause
interference with the infrared
system.
Please observe the minimum
interval of 1 minute between
two readings - a 3 min delay is
advised.
Please respect the measuring
distance (between 2 and 5 cm /
0.79 - 1.97 in).