7. Maintenance and cleaning
To replace/clean the O-ring at the front, proceed as follows:
1. Open the screw
18
2. Turn the hand test pump (with the opening
case with the palm of the hand.
The non-return valve, consisting of the valve plunger, O-ring and spring, falls out of
the opening.
The non-return valve consists of very small components! To avoid any loss,
place a container under the test pump!
3. Clean the O-ring at the valve plunger or, if worn, replace it.
4. Re-insert the non-return valve consisting of the valve plunger, O-ring and spring.
CAUTION!
Damage to the hand test pump due to incorrect mounting
Incorrect insertion of the non-return valve may lead to damage of the test
pump!
Use tweezers when inserting the valve plunger/spring!
Lightly tap the side of the pump case until the valve plunger/spring is
seated correctly.
Pay attention to the correct order of the components (see Fig. 1 "Inserting
the non-return valve")!
Do not interchange the screws from both non-return valves!
5. Re-close the opening
WIKA operating instructions, model CPP7-H
at the rear of the pump.
with the screw
18
downwards) and gently tap the pump
18
.
15
EN
18
21