Medela Invia Liberty Instrucciones De Uso página 23

Ocultar thumbs Ver también para Invia Liberty:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

y
Dressing:
- Check dressing for air leakage. Press firmly around the edges of the dressing, around
the drain tube or on the Transfer Pad.
- Apply some additional film dressing to seal the leaking area.
Connectors:
- Ensure that the tube connected to the dressing is connected properly to the Canister
tube.
- Ensure that the canister tube is inserted straight into the pump.
Canister:
- Ensure that the canister is properly inserted, release the canister and reposition.
- Ensure that the O-ring / gasket, placed beside the canister tubing on the pump is not
missing. Additional O-ring is available via Medela representative.
Recharge the battery either by placing the Invia Liberty Pump in the Docking Station
or plug in the charger to the electrical outlet port on the pump.
Remaining time of battery is approximately 15 minutes.
Change the canister, see chapter "Change Invia Liberty Canister and Invia Liberty
Tubing".
Change canister, see chapter "Change Invia Liberty Canister and Invia Liberty Tubing"
instr_Invia_liberty_patient_EN_ES_FR.indd 23
Remarks / potential cause of fault
23
31.07.13 11:33