For all ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The [MDS3002/MDS3002LA] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [MDS3002/MDS3002LA] should assure that it is used in
such an environment.
IMMUNITY
IEC 60601test
test
Radiated RF
3 V/m 80 MHz
IEC 61000-
to 2,5 GHz
4-3
Compliance
level
level
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the
[MDS3002/MDS3002LA], including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
 
Recommended separation distance:
 
-30-
80 MHz to 800 MHz
800 MHz to 2,5 GHz