atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, pre-eclampsia, renal
diseases.
8) It is suggested that the blood pressure monitor be kept at least 30 cm away
from other wireless devices, such as WLAN unit, microwave oven, etc. It can't
be used near active HF SURGICAL EQUIPMENT and the RF shielded room of
an ME SYSTEM for magneticresonance imaging, where the intensity of EM
DISTURBANCES is high.
12. If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the
procedure of blood pressure measurement, a signal of
this condition, the Electronic Sphygmomanometer can keep function, but the
results may not be accurate, it's suggested that you consult with your physician for
accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period ≥0.14s, and the number of such pulse
takes more than 53 percentage of the total number of pulse.
13. Please do not use the cuff other than supplied by the manufacturer, otherwise it may
bring biocompatible hazard and might result in measurement error.
14.
The monitor might not meet its performance specifications or cause safety
-11-
will be displayed. Under