Cleaning
1. Make sure the instrument jaws are open.
2. Immerse the product completely in the cleaning bath, with-
out the product touching other parts in the bath. Observe the
recommended immersion time.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Move the movable parts back and forth (at least 5 times).
5. Brush the inside of the shaft tube with a round brush (at
least 5 times).
6. Rinse the surfaces with sterile, deionized water (at least
1 minute).
7. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
8. Repeat the preceding cleaning steps if the product or used
rinse water still contains visible contamination.
Disinfection
1. Make sure the instrument jaws are open.
2. Immerse the product completely in the disinfectant bath,
without the product touching other parts in the bath. Ob-
serve the recommended immersion time.
3. Move the movable parts back and forth (at least 5 times).
4. Completely fill a disposable syringe with disinfecting solution
and rinse the lumina (at least 5 times).
5. Rinse the surfaces with sterile, deionized water (at least
1 minute).
6. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
7. Disconnect the irrigation tube from the shaft tube.
8. Dry the product (including the lumina) with filtered com-
pressed air.