ENGLISH
06
The electrical system must comply with the regulations in force. Verify that the
system is appropriate for the characteristics of the device.
Make sure that the measured value of the mains voltage corresponds to that
specified on the rating plate of the device.
Place the device near a power outlet, in a dry place, protected from impact or
liquid spillage.
Place the device on a perfectly horizontal surface, stable and suitable for the
weight of the appliance
The device must be moved only using the appropriate side handles.
DRAIN VALVE CONNECTOR (Only for versions with drain)
Remove the cap on the exhaust (TS) (unscrew the cap counter-clockwise).
Screw the supplied drain (RS) (turn clockwise).
Connect a drain hose to the desired length (accessory not supplied) suitable for fitting
(fitting diameter 10-12 mm).
Tighten the hose with a hose clamp (accessory not included).
RS
10
INSTALLATION
TS
VL
SA