LaMotte WaterLink SpinTouch 3580 Guia De Inicio Rapido página 3

Ocultar thumbs Ver también para WaterLink SpinTouch 3580:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 4
DOS & DON'TS
¬ Hold syringe vertically when filling disks.
¬ Do not fill disk while in the meter. Fill disk on clean, dry surface.
¬ The disk should not contain any large air bubbles. Air bubbles will
result in erroneous results.
¬ Only the Universal Disk Cover (Code 1719) can be used with the
WaterLink Spin Touch
.
®
¬ Empty syringe of previous sample before filling with next sample.
¬ Remove filled disk from meter after testing.
¬ Store disks at 70°-80°F/21°-27°C.
¬ Keep the chamber clean and dry. Gently swab LED and photodiode
lenses located around the hub with a cotton swab dampened with
streak-free window cleaner. Do not use alcohol. It will leave a thin
film over the lenses when dry.
¬ Do not store the disk cover on the hub.
FILLING: Multiple-Use Disks
When the syringe is placed in the water
sample, and the plunger is pulled all the
way up, the syringe will hold more than
enough water sample to adequately fill
the disk. Hold the syringe vertically and
insert the tip into the fill hole in the disk.
Press the plunger slowly and smoothly
to fill the disk. Sections do not have to be
filled in order. The meter will detect which
section was most recently filled.
2. Do not overfill the disk. If the disk is
overfilled, sample water will flow out
of the vent hole in the center of the
disk. If this happens sample water
may flow into a reagent chamber
prematurely and results will not be
accurate. Begin a new test in the next
chamber.
4. Do not introduce air bubbles into
the disk. The reagent chambers
will not fill entirely and results will
be inaccurate. As soon as a bubble
starts to form, pull back on the
plunger to draw the bubble out of the
disk. Begin the filling process again.
6. Disks should be filled and used within
10 minutes. They cannot be filled
ahead of time.
Disk US Patent No. 8.734.734
FCl US Patent No. 8.987.000
FCl EU Patent No. EP2784503 A1
TCl US Patent No. 8.993.337
Over Flow
Chamber
Blank
Chamber
Overfilled
(Dry SpinDisk before placing in meter)
Air Bubble
A4 manual QRC
Scan to access, download, and print the SpinTouch Manual.
Water Testing Leader Since 1919
LaMotte, WaterLink, Spin Touch are ® registered trademarks of LaMotte Company
© 2023 LaMotte Company, All Rights Reserved.
3
CONNECTING VIA BLUETOOTH
The WaterLink Spin Touch is capable of connecting to a Bluetooth
enabled device, such as a phone or tablet. The Spin Touch is also
compatible with a BLE Mobile Printer (5-0067). Other Bluetooth
printers are not supported.
Transferring results over Bluetooth to a mobile device requires that
a mobile app is installed on the device and an active account for the
associated software program. For example, with an active WaterLink
Solutions
account and the WaterLink Solutions mobile app on a
phone or tablet, results can be transferred from the meter into the
application. Mobile apps for LaMotte software products are available on
iTunes® (iOS® devices) and Google Play (for Android
For detailed information read the Spin Touch manual. Go to
softwarecenter.lamotte.com
to learn more about LaMotte software
products.
1. Fill any unused section of the disk
using slow, even pressure. Add
the sample water until the Blank
Chamber is full and the Over Flow
Chamber is partially filled.
3. Do not under fill the disk. If the disk
is under filled, the blank chamber will
not be full. The reagent chambers
will not fill entirely and results will be
inaccurate.
5. Wet disks should be dried thoroughly
with a lint free wipe. Handle the disk
by the edges. Between tests store
the disk on a flat surface. Do not
agitate a partially filled disk to avoid
leaking and contamination. Partially
used disks should be discarded at
the end of the day.
devices).
TM
Over Flow
Chamber
Blank
Chamber
Correct Fill
Blank
Chamber
Underfilled
Certified to
NSF/ANSI/CAN 5
0
3580-QG-A4 | EN | 04.20.2023
loading

Este manual también es adecuado para:

Waterlink spintouch 3581Waterlink spintouch 3582