9.3.3 Stereomicroscope with projection lens
1. Use a low power magnification and focus the specimen.
2. Reach the highest magnification available using the
zoom knob and then focus the specimen again.
3. Turn on the live-view on the camera, without changing
the focus on the microscope.
4. While observing the image on the screen/monitor, un-
screw the fixing knob ① of the trinocular port (Fig. 12)
and move up or down the movable part of the adapter
② until the focus is achieved on the camera. (Fig. 13)
5. Lock the fixing knob ①. (Fig. 12)
The proper parfocality adjustment is obtained when the
same focus is reached while looking into the eyepieces
and on the screen/monitor.
②
8
①
F ig. 12
F
ig. 12
F ig. 13
F
ig. 13