5 Solución de problemas
Código
de error
0x940E
0x940F
0x9410
0x9411
236
Serie CL de Nexera
Mensaje de error
Superior: pantalla del panel
de control
Inferior: pantallas de la
estación de trabajo y del
controlador
WARNING
LOW SET TEMP
Column oven: The
temperature set currently is
too low compared to the room
temperature. Change setting
temperature because it may
not reach set temperature.
ERROR
WRITE TO CMD
Column oven: Writing to
Column Management Device
(CMD) failed. Do not remove
CMDs during analysis or
writing.
FATAL
AUX HEATER
Column oven: The
temperature control was
stopped because the auxiliary
heater did not heat up.
ERROR
DOOR SENSOR
Column oven: The door is not
completely closed for a long
period of time. Check that the
door is closed securely.
Causa principal
Compruebe la configuración de la
temperatura del horno de columnas
en los ajustes de parámetros del
La configuración es
instrumento. Si está a más de 12 °C
demasiado baja en
de diferencia con respecto a la
relación con la
temperatura ambiente, cambie la
temperatura ambiente.
configuración para que se
encuentre dentro de los 12 °C de
diferencia con respecto a la
temperatura ambiente.
No desconecte el CMD durante un
Error al escribir en el
análisis o durante la escritura.
dispositivo de gestión
Si sospecha un fallo del CMD,
de columnas (CMD).
sustituya el CMD.
El calefactor auxiliar no
APAGUE el instrumento y póngase
funciona
en contacto con un representante
correctamente.
de ventas o servicio de Shimadzu.
Asegúrese de que la puerta esté
bien cerrada. En el caso de que esté
La puerta del horno de
completamente cerrada, podría ser
columnas se ha dejado
un fallo del sensor de la puerta.
entreabierta durante
APAGUE el instrumento y póngase
mucho tiempo.
en contacto con un representante
de ventas o servicio de Shimadzu.
Contramedida