Only then does the charging process for the battery in
the rear charging bay start.
The charger monitors the battery status and indicates
the charge state via two LED indicators.
LED indicators:
• Blinking green: battery charging
• Illuminated green = battery is completely charged
• Blinking red = battery or charging station defective
Have the manufacturer repair the product.
2.6.2 Charging time
Charging is stopped as soon as the maximum charging
time is reached.
The charging time is approx. 240 min.
3.
Preparation and setup
Non-compliance with the following instructions will
preclude all responsibility and liability in this respect
on the part of Aesculap.
When setting up and operating the product, adhere
to
– national regulations for installation and opera-
tion,
– national regulations on fire and explosion protec-
tion.
– operating advisories according to IEC-/VDE regu-
lations.
Fire and explosion hazard!
Do not operate the product in
explosion-hazard areas.
DANGER
Malfunction of, or damage to the
product caused by improper setup!
Do not expose the product to
direct sunlight or moisture.
CAUTION
Do not cover the ventilation
slots of the product.
Connect the product to the
specified voltage of 100 V to
240 V/50 Hz to 60 Hz.
Ensure that the mains power
socket used for supplying the
product is freely accessible.
5