All connections and disconnections must be performed by qualified and
▪
appropriately trained staff.
4.1
Specifications and use
4.1.1 Nurse Call cable specification
Wires length = 5 cm Shrinkable tube = 2 cm
The unterminated side is composed of 3 wires, connected to Nurse Call signalling
system:
▪
R: Stand-off contact
▪
T: Working contact
C: Common contact
▪
4.2
Connection to the holder
▪
Connect the terminated Nurse Call cable to holder Nurse Call connector.
▪
Pinch the cable into the slot provided for this purpose.
▪
Connect the Nurse Call cable unterminated side to the hospital Nurse Call system
as described below.
▪
Check that the Nurse Call system is functional by generating alarms, Refer to
Nurse Call quick check protocol
connected Nurse Call system.
▪
To disconnect, unplug the Nurse Call cable.
4.2.1 Nurse Call system electrical schematics
Without pump alarms:
System operation
Pump is running without any alarm
Pump is OFF
No pump is installed on the holder
Pump is booting
10
. Ensure the pump alarm is transmitted on the
Contact
C-T
Electrical continuity
C-COM-NO-T