4.2
Mounting the EIB 5000
4.2.1
EIB 5181
Variant 1
Figure 5: Mounting variant 1 with adapter
Mount the EIB 5181 with the top hat rail adapter to a
top hat rail (depth: min. 15 mm)
Recommended: EN 50022, 35 mm x 15 mm
Connect the protective ground to the appropriate M5
thread of the housing
The cross section for connecting the
protective conductor should be at least
twice the cross section of the lines for the
temperature measurement; a minimum cross
section of 2.5 mm² is required, however.
48
Variant 2
2
1
Figure 6: Mounting variant 2 without adapter
Remove the top hat rail adapter by loosening the three
screws (1)
Use two M3 screws to fasten the EIB 5181
Connect the protective ground to the appropriate M5
thread of the housing
The cross section for connecting the
protective conductor should be at least
twice the cross section of the lines for the
temperature measurement; a minimum cross
section of 2.5 mm² is required, however.