Operation
6.17.1

STM function

56 / 82
delay temp alarm: The alarm is activated with a delay (of 1 min,
2 min, 5 min, 10 min, or 15 min) after the permitted temperature
limit is exceeded.
delay door alarm: The door alarm is be activated when the door
remains open for the time selected by the user (30 s, 1 min, 2
min, 5 min, or 10 min).
Turning the STM function On or Off (to read the description of this
function, go to Section...)
Fig. 52: Alarms
Confirms changes
Cancels the entered changes
In the field "lower alarm temp" you can set a value
anywhere from -0.5°C to -5°C, and in the field
"upper alarm temp" a value anywhere from +0.5°C
to +5°C.
The Smart Temperature Monitor (STM) function informs the user
if any problems incurred with reaching or maintaining the set tem-
perature. The user can activate/deactivate the function. If the STM
function is ON, the STM icon appears on the screen next to the
temperature of the main sensor.
02.2023 / Rev. --
AB69601en