Loading And Overloading Of Rotors; Removing The Rotor; Power Switch; Lid Control - OHAUS Frontier FC5706 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

EN-6
ATTENTION!
Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or mechanical
damage.
Do not operate with extremely corrosive substances, which could damage the rotor and buckets.
In case of any questions, please contact the manufacturer!

3.1.4 Loading and overloading of rotors

All approved rotors are listed with their maximum speed and maximum filling weight in "table 2 permissible net
weight" (See APPENDIX).
The maximum load permitted for a rotor, which is determined by the manufacturer, as well as the maximum speed
allowed for this rotor (See label on rotor), must not be exceeded. The liquids the rotors are loaded with should have
a maximum homogeneous density of 1.2 g/ml or less when the rotor is running at maximum speed.
In order to spin liquids with a higher density, the speed has to be reduced according to the following formula:
Reduced speed n
=
red
Example:
1,2
n
=
x 4,000 = 3,360 rpm
red
1,7
In case of any questions, please contact the manufacturer!

3.1.5 Removing the rotor

Untighten the rotor fixing nut completely (screw over the stiff point) and lift the rotor vertically out of the centrifuge.
(See figure 12)
3.2

Power switch

The power switch is located on the bottom left side of the unit (see figure 17).
Figure. 17: Power switch
3.3

Lid control

3.3.1 Lid open

After the run, when closing the lid of the centrifuge, in the display "RPM | RCF" (A-1) the word "close" (M1)
appears (refer to figure 18 below).
If there is a rotor in the centrifuge, additional the word "rotor"(M3) appears, as well as the code number of the
respective rotor which is in the centrifuge system "77" (M4). If there is no rotor in the centrifuge, the word "rotor"
(M3) flashes and additional the word "no" (M4) appears. By pressing the key "Door Open" (7) you can release the
lid of the centrifuge. As soon as the electromagnetic lid is completely released, the word "open" (M2) appears.
Now you can open the lid of the centrifuge.
1,2
x max. speed (n
ℎ����ℎ���� ��������������
Frontier
) of the rotor
max
TM
Centrifuge FC5706
Tabla de contenido
loading

Tabla de contenido