True RMS Thermal MultiMeter / English
4.13.AC+DC Current Measurements
CAUTION: Do not make 10A current measurements for longer than 30
seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test
leads.
• Insert the black test lead banana plug into the negative COM jack.
• For current measurements up to 6000μA AC+DC, set the function
switch to the yellow μA position and insert the red test lead banana
plug into the μA/mA jack.
• For current measurements up to 600mA AC+DC, set the function
switch to the yellow mA position and insert the red test lead banana
plug into the μA/mA jack.
• For current measurements up to 10A AC+DC, set the function
switch to the yellow 10A position and insert the red test lead banana
plug into the 10A jack.
• Press the MODE button to indicate " " on the display.
• Read the current in the display.
4.14.Using RANGE
Press the RANGE key to activate the manual mode and to disable the
Autorange function. The message "Manual Range" appears on the
upper left part of the display instead of "Auto Range". In manual mode,
press the RANGE key to change measuring range: the relevant decimal
point will change its position. The RANGE key is not active in positions
o
o
Temp
C
F 10A
In Autorange mode, the instrument selects
the most appropriate ratio for carrying out measurement. If a reading
is higher than the maximum measurable value, the indication "O.L"
appears on the display. Press and hold the RANGE key for more than 1
second to exit the manual mode and restore the Autorange mode.
12
4.15.Hold Mode
To freeze the display for any function,press key HOLD.
And again press key HOLD to release freeze.
4.16.Capturing Minimum and Maximum Values
The MAX MIN Record mode captures minimum, and maximum input
values.
When the input goes below the recorded minimum value or above the
recorded maximum value, the Meter beeps and records the new value.
This mode is for capturing intermittent readings, recording minimum
and maximum readings unattended, or recording readings while
equipment operation precludes watching the Meter.
To activate the MAX MIN mode, press soft key labeled . If the Meter
is already in MAX MIN function, press
MAX MIN function.
4.17.Relative Values
To activate the relativemode, press the soft key labeled .
If the Meter is already in the relative function, press
Meter to turn off relative.