Shorten Suspension Cable; How To Proceed If A Repair Is Necessary; Dismount; Dismounting Steps - Vega WELL S 51 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6 Dismount

Fehlercode
Ursache
4 ... 20 mA
No atmosphe-
signal not
ric pressure
stable
compensation
4 ... 20 mA
Wrong
signal missing
connection to
voltage supply
No power
supply
Operating
voltage too
low or load
resistance too
high
In Ex applications, the regulations for the
wiring of intrinsically safe circuits must
be observed.
5.3

Shorten suspension cable

Shorten the suspension cable individually.
Proceed as follows:
1. Remove the filter adapter from the capillary
line
2. Cut the suspension cable to the requested
length with side cutters
Caution:
Do not squeeze the capillary cable, as
this will impair the pressure compensati-
on. If necessary, rework the capillary with
a sharp knife.
3. Remove approx. 10 cm of the cable mantle,
strip off approx. 1 cm of insulation from the
ends of the wires
4. Insert the filter adapter
The work steps are finished.
5.4
How to proceed if a repair is
necessary
You can find an instrument return form as well
as detailed information about the procedure in
the download area of our homepage. By doing
this you help us carry out the repair quickly and
without having to call back for needed informa-
20
Beseitigung
Check the capillariy and cut it clean
Check the pressure compensation and clean the filter element, if
necessary
Check connection according to chapter " Connection steps" and
if necessary, correct according to chapter " Wiring plan"
Check cables for breaks; repair if necessary
Check, adapt if necessary
tion.
In case of repair, proceed as follows:
6
6.1
Take note of chapters " Mounting" and " Con-
necting to voltage supply" and carry out the
listed steps in reverse order.
6.2
Pass the instrument on to a specialised recy-
cling company and do not use the municipal
collecting points.
Remove any batteries in advance, if they can be
Print and fill out one form per instrument
Clean the instrument and pack it damage-
proof
Attach the completed form and, if need
be, also a safety data sheet outside on the
packaging
Ask the agency serving you to get the
address for the return shipment. You can find
the agency on our homepage.
Dismount

Dismounting steps

Warning:
Before dismounting, be aware of dan-
gerous process conditions such as e.g.
pressure in the vessel or pipeline, high
temperatures, corrosive or toxic media
etc.

Disposal

VEGAWELL S 51 •