Change factory settings
5. If no scale factor changes are necessary, go to save changes (see point 9.4).
Changing the unit of measurement from metric to English or from English to metric
clears the resettable total and accumulated total.
9.3

Change scale factor

Changing the scale factor changes the accuracy of the meter, potentially causing it to
overfill or under fill. This has the potential to cause a mechanical breakdown.
Scale factor digits
3. Press the
4. Repeat steps 2 and 3 for all five digits in scale factor.
9.4

Save changes

To save changes and exit the programming mode:
1.
Push and hold at the same time the
2.
The display will flash 3 times and go blank.
3.
Press the
9.5

Verify changes

1. Verify unit of measure is correct.
2. Push and hold the total button and auto button together, to verify that the scale
factor is correct.
1. Press the
factor digits.
2. Press the
number.
NOTE: All digits can be scrolled between 0 and 9
except the first. It can only be scrolled from 0 to 1 or
from 1 to 0.
button to advance to the next number in scale factor.
button and the display will turn back on.
LM_OGP2Adv_BA_89_1008
button to advance through the scale
button to change the selected
and
buttons.